VQC Processing and Software Realignment

1.) Set the system into PET mode by clicking the [PET] icon on the Image Monitor desktop area.
The Scan Monitor displays the PET main screen.
2.) Click the [Full QA] button on the PET main screen.
3.) Select Volumetric QC from the drop-down menu. The Volumetric QC panel shown in
Figure 8-6
4.) Refer to
The Images Database Navigator displays showing the Patient List.
Page 248
6
displays.

Figure 8-6: Initial Volumetric QC Panel

Figure
8-7. Click [Load] to load the image sets that represent the CT and PET planes.

Figure 8-7: Images Database Navigator

Chapter 8 - System QC, Calibration and Functional Tests