Note:
• Do not use the Scan Stop
• If it is necessary to stop gantry rotation immediately, use the E-stop shown in
Figure
rotation.
2.2.3.3
Table Latch and Tape Switches
Pressure sensitive tape switches are located on both sides of the cradle and base. The purpose of
these switches is to sense obstructions during cradle movement. When activated, the switch
disables cradle drive power. The cradle unlatches when cradle drive power is removed.
In case of an emergency, a Cradle Latch button
It is a toggle switch. When engaged, it unlatches the cradle, but does not remove power to the
cradle's drive. To relatch the cradle, simply press the cradle latch button again.
Note:
Never use the Cradle Latch/Unlatch button (shown in
cradle's positioning buttons instead. Sudden movement of an unlatched cradle can cause the
system to lose track of positioning information, especially during use of an injector.
6
2-6. The scan abort switch only terminates X-ray generation and does not stop gantry
Tape Switches

Figure 2-8: Table Tape Switches

Figure 2-9: Cradle Latch Button

Chapter 2 - Discovery LS System Safety Information
button on the console or the gantry control panels.
TOUCH SENSOR.
(RIGHT)
is provided on the gantry control panel.
Figure
2-9) to position patients. Use the
D
LS S
S
ISCOVERY
YSTEM
ERVICE
TOUCH SENSOR.
(RIGHT)
TOUCH SENSOR. (LEFT)
Cradle Latch/
Unlatch Button
M
ANUAL
Page 47