6.2.6.38

Adjust the Table Latch

Follow the procedure in this section to adjust the CT-PET Position assembly and Table Latch.
1.) Push the table into the CT position. Lower the Table Latch bar and pin into place.
(Figure 6-103
2.) Refer to
the solenoid latch arm and create a gap of about 3 to 3.5 mm between the bottom of the detent
and the bottom of the Table Latch Arm. (The latch should not allow forward or backward table
movement at this time.)
Page 182
6
shows the Table Latch pin in place.)
Figure 6-101
and
Figure
3 mm Allen Wrench
Bottom of Table latch should clear the detent by 3 mm.

Figure 6-101: Adjust CT Position Assembly to Clear Table Latch Arm

Use the 3 mm Allen wrench as a gauge for 3 to 3.5 mm gap here.

Figure 6-102: Correctly Adjusted Position Latch Arm

Chapter 6 - Table
6-102. Use a 3 mm Allen wrench to adjust the screw to lower
D
LS S
S
ISCOVERY
YSTEM
ERVICE
Solenoid
Latch Arm
M
ANUAL