Attach the Top Cap Assembly to the Secondary Base
Refer to
secondary base, and rests on flanges attached to the side rails. When the table moves back to the
CT detent, the metal Table Top Cover, P/N 2333078, slides beneath the Top Cap Assembly.
The Top Cap attaches with a bracket to the foot end of the secondary base.
1.) Use a 5 mm Allen wrench to fasten the Top Cap bracket into place with two sets of the following
M5 hardware:
• M5 x 16 socket head cap screw (SHCS), P/N 46-328417P505
• M5 lock washer, P/N 46-328432P8
• M5 flat washer, P/N 2236649
6
Figure
6-134. The Top Cap Assembly, P/N 2320161, attaches to the foot end of the
Figure 6-134: Table Top Cap Assembly (P/N 2320161)
Chapter 6 - Table
D
ISCOVERY
Use 5 mm Allen wrench.
Use 8mm Allen wrench.