2.4 Emergency procedures

This section describes how to do an emergency shutdown of the Typhoon instrument.
This section also describes the result in the event of power failure.
Emergency procedures
In an emergency situation, do as follows to stop the run:
Step
1
2
Power failure
Power failure to...
Typhoon instrument
Computer
Typhoon Operating Instructions 28-9623-56 AA
Description
This symbol indicates that the waste of electrical and electronic
equipment must not be disposed as unsorted municipal waste and
must be collected separately. Please contact an authorized
representative of the manufacturer for information concerning the
decommissioning of equipment.
This symbol indicates that the product contains hazardous materials
in excess of the limits established by the Chinese standard SJ/
T11363-2006 Requirements for Concentration Limits for Certain
Hazardous Substances in Electronics.
Action
In the Typhoon Scanner Control software, click CANCEL SCAN to abort
the scan.
If required, press the power switch to the OFF position to turn the
instrument off.
will result in...
The run is interrupted immediately, in an undefined
state.
The data collected up to the time of the power failure
is available in the file created when starting the scan.
The computer shuts down in an undefined state.
The run continues, but data cannot be saved.
Safety instructions 2
Emergency procedures 2.4
21