6.2.5.4

Remove the Pedal Assembly

1.) Refer to
that fasten the pedal to the side of the table base.
2.) Remove the front left table side cover from the scissor section of the table.
a.) Use an 1/8 inch Allen wrench to remove the two screws and ground wire that fasten the
b.) If necessary, move the table to access the front switch rail screws.
6.2.5.5

Remove the PET and CT Position Switches and Switch Rail

Locate the CT and PET position limit switches (mounted to the switch rail and attached to the left
side of the secondary base when facing the gantry). Move the table, as needed, to access the
Position switch hardware and components. The cables may be cut and the switches and switch rail
removed as a single assembly.
1.) Refer to
switches before removing the CT position switch from the table.
2.) Remove the retaining rings from PET position switches as well.
Page 124
6
Figure
6-15. Use a 6 mm Allen wrench to remove the two socket-head cap screws
Socket-Head
Cap Screws

Figure 6-15: Remove Pedal from Table Rail

front left table side cover into place. (Only need to remove one cover.)
Figure 6-16
and
Figure
Retaining Rings

Figure 6-16: PET Position Switches (Shown with Pedal Assembly Attached to Rail)

Chapter 6 - Table
6-17. Use a wrench to loosen the two retaining rings from the
D
LS S
S
ISCOVERY
YSTEM
ERVICE
M
ANUAL