Do not install the plate backward.
3.) Orient the plate as shown in
• The edge of the plate with two holes faces the rear of the table.
• The edge of the plate with one hole faces the front of the table.
4.) Fasten the plate into place with the corresponding hardware.
6.2.6.27
Install the PET-CT Position Latch Assembly
The CT-PET Position Latch Assembly contains the optical sensor that detects the PET and CT
flags, the latching mechanism that holds the table in the PET and CT position, and a solenoid latch
that holds the table in the CT (patient load) position.
1.) Refer to
Page 166
6
To Clutch
Motor
Figure 6-77: CT-PET Mounting Plate in Correct Orientation
Figure
6-78.
Connect Cable P/N 2338086 tabs to provide AC power to the solenoid.
Solenoid-controlled latch holds the table in the CT (patient load) position.
Figure 6-78: CT-PET Position Latch Assembly, P/N 2340721
Chapter 6 - Table
Figure
6-77.
Table Rail
Rolls into CT and PET detent to hold table position.