6.2.6.37
Install the Table Latch
1.) Refer to
During normal operation, the latch holds the table in the patient loading (CT) position.
2.) Refer to
is perpendicular to the bracket.
• If necessary, use a 1.5mm allen wrench to adjust the latch bar.
• If available, apply Loctite® 242 to the set screw.
NOTICE
The latch pin limits the amount of adjustment to the set screw. For best results, insert the
latch pin before adjusting the set screw.
NOTICE
The Table Rail revision determines the length of the fastening hardware.
Page 180
6
Figure
6-98. Locate and remove the Table Latch Assembly from the upgrade kit.
Table Latch
Figure 6-98: Table Latch Assembly
Figure
6-99. The latch assembly should arrive with the set screw adjusted so the bar
Use a 1.5 mm Allen wrench to adjust set screw to raise and
lower the bar (should be perpendicular to the bracket).
Figure 6-99: Adjust the Set Screw for 90 Degrees
Chapter 6 - Table
D
LS S
S
ISCOVERY
YSTEM
ERVICE
Latch Pin
M