or near the eye should be carefully screened before undergoing an MR exam.
Certain transdermal patches may cause burns to the underlying skin due to absorption of RF energy.
The supplier of the patches should be consulted or the patch should be removed to avoid burns. A
new patch should be applied after the examination.
Patients shall be supervised at all times during scans.
The use of auxiliary equipment that has not been specifically tested and approved for use in the
environment of the MR equipment may result in burns or other injuries to the patient. Auxiliary
devices labeled as compatible with MR equipment or MR systems may be capable of causing injury if
this operator manual is not followed.
Equipment not suitable for use in the presence of a flammable anesthetic mixture with air or with
oxygen or nitrous oxide.
The following are coil specific warnings statements
This Product contains chemicals, including lead, known to the State of California to cause birth
defects or other reproductive harm. Wash hands after handling.
Warning, Burn Hazard. Remove this coil from the MR System when not plugged in. Failure to do
so may result in injury and/or damage to the coil
Do not cross or loop cables. Arcing and patient burns could result. Route cables out of the
magnet so that they do not touch the patient.
Visually inspect the cable insulator jackets, strain reliefs, and connector boxes before each use. If the
insulation is broken, or if the cable is frayed, immediately discontinue use of the device.
Use only approved accessories.
Do not lift or carry coil by bail (latch).
Ensure that the patient does not form a loop with any body parts. Do not allow patients to touch any
part of the right hand or arm to any part of the left hand or arm. The loop formed by doing so could
cause an RF burn at the point of contact. Use pads to ensure the patient's hands do not touch any
metal features of the patient table or cradle. Be sure to educate the patient accordingly and check
the patient's position immediately before the scan.

Emergency procedures

In the unlikely event that a coil creates smoke, sparks or makes an unusually loud noise, or if the patient
requires emergency assistance:
Stop the scan if one is in progress.
Remove the patient from the scan room if medical treatment is needed.

Technical considerations

The coil and accessories require special conditions regarding electromagnetic compatibility. The coil