Figure 8-5: Selecting The Protocol - GE Discovery LS System Service Manual

Table of Contents

Advertisement

GE
HEALTHCARE
D
2307224-100, R
IRECTION
EVISION
4.) Refer to
the Volumetric QC protocol from the drop-down menu.
5.) When the CT VQC Protocol is displayed, verify that the PET Protocol icon is enabled and PET
VQC Protocol is selected.
6.) Verify that Auto Transfer (to the PET OWS station) is On.
7.) Verify that the VQC phantom is properly positioned and oriented. (Refer to
8.) Refer to
scan takes a few seconds.)
9.) When the CT scan is completed, click the [Pet Series] icon on the Scan Monitor.
10.) Click [Accept Setup], and perform PET Transmission VQC acquisition.
• Confirm that there is no emission activity by clicking [OK] in the dialog box. The scan takes
• If the count rate is less than 1000 Kcps, increase the scan time to 10 minutes.
11.) When the acquisition is finished, select Quit Screen from the Options menu to close the PET
Acquisition screen.
12.) Select Recon from the [Screens] drop-down menu.
13.) When the Recon panel displays, click [Next Recon].
14.) When the Navigator panel opens, select the scan that was just performed and click [OK].
15.) Click [Defaults Protocols], and select VQC Reconstruction Protocol.
16.) Click [Submit to Top].
17.) After reconstruction is complete, select Quit Screen from the Options menu to close the
Reconstruction screen.
6
Figure
8-5. Click below the patient's legs in the Protocol Selection area, and choose

Figure 8-5: Selecting the Protocol

Figure
8-1. Set the <Start Landmark> on the Start Acquisition Line the scan. (The
5 minutes (default time).
Chapter 8 - System QC, Calibration and Functional Tests
D
LS S
S
ISCOVERY
YSTEM
The Volumetric QC
protocol can be
selected by clicking
under the feet in the
picture of the patient.
Section
M
ERVICE
ANUAL
8.1.1.3.)
Page 247