Base Axis Does Not Move; Figure 12-12: Troubleshooting For Base Axis Does Not Move - GE Discovery LS System Service Manual

Table of Contents

Advertisement

GE
HEALTHCARE
D
2307224-100, R
IRECTION
EVISION
12.4.2

Base Axis Does Not Move

If the base axis does not move, perform the following checks:
Reset the E-STOP
circuitry or press the
Cradle Release button
6
Base Axis does not move
Is E-STOP circuitry
yes
activated or Cradle Release
Release the
mechanical
yes
interference
Check:
1. The path from the ELEVATION/TILT(BASE Axis ) Amplifier to
the motor include the Amplifier itself
PNs:
1. 2323724
2. 2309058
3. 2323720
4. 2323722

Figure 12-12: Troubleshooting for Base Axis Does Not Move

Chapter 12 - Troubleshooting
button pressed?
Does a mechanical interference exist?
Check the locking mechanism including the
solenoid functionality, use the PET position Jog
button on the Control Panel
End of test
D
LS S
S
ISCOVERY
YSTEM
ERVICE
no
no
Page 333