Switching Off Power To Pet Subsystems And Peripherals; Figure 3-15: Ct Computer; Figure 3-16: Pet Computer - GE Discovery LS System Service Manual

Table of Contents

Advertisement

GE
HEALTHCARE
D
2307224-100, R
IRECTION
EVISION
a.) To turn off the system, press the [Power] button on the CT computer.
b.) To restart the system, click [Restart] or press any key on the keyboard.
4.) To shutdown the PET computer, press the [Power] button on the Sun computer.
3.11.2

Switching OFF Power to PET Subsystems and Peripherals

In the current release, the system shutdown procedure shuts down the CT subsystems only. The
PET subsystems shut-down procedure is described below.
To switch OFF power to PET subsystems and peripherals:
1.) If necessary, turn OFF the PET computer power switch.
2.) Turn off all the OWS peripherals plugged into the distribution box.
3.) Refer to
6

Figure 3-15: CT Computer

Figure 3-16: PET Computer

Figure
3-17. Switch the PDU circuit breakers, located on the back side of the
Chapter 3 - DLS System Functional Description
D
ISCOVERY
Power button
Power Button
LS S
S
M
YSTEM
ERVICE
ANUAL