Refer to
present.
Room Warning Light
If a room warning light was installed and connected to the CT system correctly, the room warning
light will illuminate whenever X-rays are present, by default. The room warning light can also be
configured to illuminate whenever high voltage is present.
See the Discovery LS System Installation Manual, Direction 2317946-100 for wiring and
configuration details.
2.2.3

Emergency Switches, Buttons and Locks

WARNING
ACTIVATION OF THE CT E-OFF (EMERGENCY OFF) DOES NOT STOP MOVEMENT OR
SHUTDOWN POWER TO THE PET MOTION CONTROL SUBASSEMBLY OF THE DLS
SYSTEM. ONLY THE E-STOP (EMERGENCY STOP) BUTTON, LOCATED INSIDE THE PET
GANTRY COVERS, WILL STOP THE PET MOTION CONTROL SUBSYSTEM IMMEDIATELY
AND SHUT DOWN POWER TO THE MOTION CONTROL.
2.2.3.1

System Emergency Off (E-Off) Switch

The E-Off switch removes all power to a CT system immediately. If for any reason the operator
needs to remove all power supplied to the CT system at the main distribution panel, the E-Off
switches should be employed. Using this switch except in the case of an emergency could cause
damage to hardware. Typically, one or more E-Off switches are located in or near the operator
console or gantry. Know their locations.
6
Figure
2-4. A backlit x-ray On indicator is located on the SCIM. It illuminates when x-ray is
X-ray On Indicator

Figure 2-4: Operator Console and Gantry X-ray Exposure Warning Lights