GE
HEALTHCARE
D
2307224-100, R
IRECTION
EVISION
6.2.6.2
Determine the Right Secondary Base Rail Revision
Determine the secondary base revision by counting the number of access holes in the right rail
while facing the Gantry. All tables use the CT Drill Template.
• Seven rail access holes: Revision 0 table needs the PET Plug and CT Plug. Drill four PET
• Five rail access holes: Revision 1 or 2 table needs the PET Plug. Drill four PET holes and
• Four rail access holes: For Revision 3 table, drill one CT hole. (May need to drill four PET
6
holes and nine CT holes.
six CT holes.
holes if the distance between the CT scan plane and PET scan plane falls outside the
specified range.)
Access Hole
Figure 6-28: Determine the Right Secondary Base Rail Revision
Chapter 6 - Table
Revision 0 rail has seven access holes.
PET Plug
CT Plug
Revision 1 rail has five access holes.
PET Plug
Revision 2 rail has five access holes.
Revision 3 rail has four access holes.
D
LS S
S
ISCOVERY
YSTEM
ERVICE
Page 133
M
ANUAL