X-Ray/key Switch and Emergency Switch

Key switch is used to control the unauthorized usage of X-ray exposure and vertical lift motor.
Signals from Key switch to Mainframe Control Board:
Name
Description
XRAYOFF
The termination which
indicates lift column enable
and x-ray disable
ON
The termination which
indicates lift column and x-
ray are both enable
COMMON
Common termination
X-rays can be enabled via the footswitch or the Hand switch. Signals from footswitch and Hand switch to MCB
are listed in the following charts. Detailed interconnect information, please refer to chapter 2.2.3
Signals from Hand Switch to MCB:
Name
HAND_FLU
Hand switch termination
HAND_HLF_SECURITY Hand security line
termination
HAND_HLF
High Level Fluoro hand
switch termination
HAND
High Level Fluoro hand
_HLU_SECURITY
security line termination
HAND_COM
Common termination
Signals from Footswitch to MCB:
Name
FOOT_FLU
FOOT_FLU_SECURITY
FOOT_ HLF
FOOT_ HLF _SECURITY
FOOT_BUTTON1
Attribution
Passive
Passive
Passive
Description
Description
Attribution
Foot switch
Passive
termination
Foot security
Passive
line
termination
High Level
Passive
Fluoro foot
switch
termination
High Level
Passive
Fluoro foot
security line
termination
Button 1
Passive
Scale
Comment
Open/Close
XRAYOFF & COMMON is connected means
XRAYOFF switch has pressed, which represents lift
column function has been selected and exposure
enable has not been selected.
Open/Close
ON & COMMON is connected means ON switch has
pressed, which represent both lift column and
exposure enable function have been selected.
-
-
Attribution
Scale
Passive
Open/Close J1-B28
Passive
Open/Close J1-C27
Passive
Open/Close J1-B7
Passive
Open/Close J1-A22
Passive
-
Scale
Pin
Open/Close
J1-A27
Open/Close
J1-C28
Open/Close
J1-A14
Open/Close
J1-C31
Open/Close
J1-A20
System Overview
Pin
Comment
HAND_FLU & HAND_COM is
connected means
HAND_FLU button has
pressed
-
HAND_HLF & HAND_COM is
connected means
HAND_HLF button has been
pressed
-
J1-C15
-
Comment
FOOT_FLU & FOOT_COM is connected
means FOOT_FLU button has been
pressed
-
FOOT_HLF & FOOT_COM is connected
means FOOT_HLF button has been
pressed
-
FOOT_BUTTON1 & FOOT_COM is
2-61