12.3.2

Base Axis Does Not Move

For flowchart see
If the base axis does not move, perform the following checks:
1.) Check if a mechanical interference exists.
2.) If the Table Base Axis does not move out of the CT position, check:
a.) If the solenoid of the Positioning Unit is faulty by pressing the PET Position Jog button on
b.) Check if the Solenoid pulls the locking mechanism down, releasing the CT Position Latch.
6
Figure
12-10.
the control panel, and
* If the locking mechanism does not release from the CT Position latch, go to
on page
326.
* If the locking mechanism releases from the CT Position Latch, go to
327.

Figure 12-6: Solenoid, Locking Mechanism and CT Position Latch

Chapter 12 - Troubleshooting
Positioning Unit
Secondary Base
D
LS S
S
ISCOVERY
YSTEM
ERVICE
Number 3.)
Number 4.) on page
CT Position Latch
Locking Mechanism
Solenoid
CT Position
Page 325