Using Vscan Air CL

Configure the device for the pre-hospital emergency environment

5-82
The following considerations should be made regarding the
configuration of the Vscan Air for use in pre-hospital emergency
environments.
The Auto-freeze time can be adjusted to keep device
scanning for a longer period if exam is expected to be
performed with no or infrequent user interface interactions
but note that freezing the image when image acquisition
finishes will reduce battery drainage and minimize device
heating when device is not being used. It can be configured
from the Menu -> Configuration -> Auto freeze time
The probe button can be configured for additional
functionality (Store or Freeze) that can be helpful during an
emergency exam to minimize interaction with the display
device. It can be configured from the Menu -> Configuration
-> Probe button. Note that accidentally pressing the button
for 3 seconds or more could lead to an unintentional shut
down of the probe.
To avoid additional delays when starting the Vscan Air probe
and app the display unit should if possible be configured to
not accept automatic app updates. If an app and/or a probe
update is to be performed the availability of the device for
scanning might be delayed by up to 3 to 5 minutes.
An internet connection will be required for a user to sign in.
Be aware that an automatic sign out configuration setting
will give users an option to control the duration of a signed in
session. Configuration is available via Menu ->
Configuration -> Automatic sign out. The default value is 24
hours.
Vscan Air
Direction GP092020-1EN Rev 18
User Manual