3.
Fit the following new seals to the source enclosure:
•
Nebulizer gas seal
•
Desolvation gas seal
Requirement:
groove as shown.
4.
Fit the following new seals to the probe adjuster assembly:
•
Probe seal
•
Nebulizer gas seal
5.
Refit the source enclosure to the instrument (see
Replacing the air filter
You must replace the air filter annually.
Required materials
•
Needle-nose pliers
•
New filter
To replace the air filter:
1.
Open the access door to the fluidics pump.
4-90
Maintenance Procedures
These seals have a special cross-section; fit them in the
Seal
Groove
page
4-9).