The IntelliStart software automatically tunes and mass calibrates the mass
spectrometer and displays performance readbacks. Integrated with
MassLynx™ mass spectrometry software and Instrument Console software,
IntelliStart software enables simplified setup of the system for use in routine
analytical and open access applications (see
"Software and data system" on
page
1-5).
1
The IntelliStart Fluidics system
is built into the mass spectrometer. It
delivers sample directly to the MS probe from the LC column or from three
integral reservoirs. The integral reservoirs can also deliver sample through
direct or combined infusion so that you can optimize instrument performance
at analytical flow rates.
Use reservoir A for the calibrant solution, reservoir B for
Recommendation:
tuning compounds, and reservoir C for analyte/optimization solution.
A fourth reservoir contains solvent for automated flushing of the solvent
delivery system.
See the mass spectrometer's online Help for further details on IntelliStart
technology.

ACQUITY UPLC/MS Xevo TQ-S systems

®
The Waters Xevo TQ-S is compatible with the ACQUITY UPLC
systems; if
you are not using an ACQUITY UPLC system, refer to the documentation
relevant to your LC system (see
"Non-ACQUITY devices for use with the Xevo
TQ-S" on page
1-4).
ACQUITY UPLC system
The ACQUITY UPLC system includes a binary solvent manager, sample
manager, column heater, optional sample organizer, optional column
manager, optional detectors, and a specialized ACQUITY UPLC column.
WatersMassLynx mass spectrometry software controls the system.
For further instruction, see the ACQUITY UPLC System Operator's Guide or
Controlling Contamination in LC/MS Systems (part number 715001307). You
can find these documents on http://www.waters.com; click Services and
Support > Support.
1. In this document, the term "fluidics" is used to describe plumbing components and fluid
pathways within and between instruments and devices.
Waters Xevo TQ-S
1-3