Operating the source isolation valve

You must close the source isolation valve to isolate the source from the
instrument vacuum system for certain maintenance procedures.
Required materials
Chemical-resistant, powder-free gloves
To close the source isolation valve before starting a maintenance
procedure:
Warning:
prepared before commencing this procedure.
1.
Prepare the instrument for working on the source (see
2.
Pull the source enclosure release (located at the bottom, right-hand side)
outwards, and swing open the enclosure.
4-10
Maintenance Procedures
The source components can be contaminated with
Warning:
biohazardous and/or toxic materials. Always wear
chemical-resistant, powder-free gloves while performing this
procedure.
To avoid electric shock, ensure that the instrument is suitably
The source can be hot. To avoid burn injuries, take great
Warning:
care while working with the instrument's source enclosure open.
To avoid puncture wounds, take great care while
Warning:
working with the source enclosure open if an ESI probe is fitted.
page
4-6).