Returning The Clarity System To Use - Integra CUSA Clarity System Service And Maintenance Manual

Hide thumbs Also See for CUSA Clarity System:
Table of Contents

Advertisement

Irrigation Test
• Press Main Screen button
• Position Aspiration Mode on Constant position
• Position the Digital Tachometer in front of the cartridge door
• Select Irrigation at 2 ml/min
• Measure the speed of the pump head – record the value
o Expected value is from 2 to 8 RPM
• Press Fast Flush pedal
• Measure the speed of the pump head – record the value
o Expected value is from 34 to 47 RPM
• Record irrigation values in IT0739A CUSA Clarity GS&R End User Functional Test Datasheet
Aspiration Test
• From Main Screen, select Aspiration Mode on Constant position
• Set aspiration at 100%
• Measure the Aspiration – record the value
o Expected value is from 304 to 436 mmHg.
• Record aspiration values in IT0739A CUSA Clarity GS&R End User Functional Test Datasheet

7.10 Returning the Clarity System to Use

When functional testing is complete, accomplish the following actions to return the CUSA Clarity system to service:
• Clean the system
• Remove the maintenance label and replace with a new maintenance label
o Include the date of service on the new label.
• Include a copy of the Electrical Safety Test Datasheet IT0738A with the CUSA Clarity End User Functional Test Datasheet IT0739A.
CUSA® Clarity System Service and Maintenance Manual
Page 61 of 117
0645403-2-EN