6.0 Test conditions
The tests may be performed at the ambient temperature, humidity, atmospheric pressure and mains voltage as present at the site of testing.
7.0 Visual inspection
Visually inspect the console for signs of damage which would affect the normal operation and function of the system.
Check the following items:
1. Carefully inspect the console for scratches on front and rear component as well as the metal cover.
2. Inspect the IV Pole at the rear side for damage and check its functionality.
3. Check that all four rubber feet are in place and not damaged.
4. Inspect for cracks or damage to touch screen display.
5. Check irrigation door for damage and check it is opening and closing smoothly.
6. Check irrigation pump head and rollers for damage.
7. Check pinch valve for damage.
8. Check contamination guard for moisture or blood.
9. Check condition the AC power cord returned with the console.
10. Check connectors for damage and proper orientation.
11. Check all labels and confirm that they are in good condition, legible, and firmly attached (no curled or loose edges).
12. Inspect for scratches or damage to foot-pedal.
13. Inspect for cracks or damage to handpiece and its cord and connector if available.
14. Inspect console to cart attachment.
15. Inspect cart for damage.
a. Inspect Base Plate for cracks or damage.
b. Inspect Vacuum Canister for cracks or damage.
c. Inspect Cart Foot Switch for cracks or damage.
d. Inspect Cart Cable Peg for cracks or damage.
e. Inspect Cart Weight Cover for cracks or damage.
16. Inspect Cart Weight Cover for cracks or damage.
17. Inspect cart wheels for proper rotation and direction.
Record results in IT0739A.
8.0 Electrical Safety Tests
• Conduct Electrical Safety test according to IT0738.
• Record in IT0739A that the Electrical Safety test passed.
9.0 Functional Test

9.1 Test Setup

• Turn on the console.
• Connect the Footswitch and the CUSA® Clarity 36 kHz Handpiece.
• Connect a contamination guard on the console and connect the canister.
• Open the cartridge door and stick the reflective tape on the border of the pump head
CUSA® Clarity System Service and Maintenance Manual
Page 110 of 117
0645403-2-EN