Test Procedure IT0739
CUSA® Clarity End User Functional Test
Issued: 27 Oct 2016
1.0 Purpose
This procedure provides a method for testing functionality of the CUSA® Clarity with the minimum test equipment that can be handled by customer
2.0 Scope
This document covers the final functional test of the CUSA® Clarity Console.
3.0 Responsibility
Trained personnel shall perform this test.
Global Service & Repair is responsible for updating and maintaining this procedure.
4.0 Reference Document
CUSA® Clarity System Service and Repair Work Instruction
5.0 Equipment
Equipment
CUSA® Clarity console system (DUT)
Footswitch
CUSA® Clarity 36 kHz Handpiece
CUSA® Clarity 36 kHz Curved Ext MicroTip(Single Pack)
CUSA® Clarity Quick Connect Tubing Set(Single pack)
Contamination guard
CUSA® Clarity 36 kHz Sterilizable Torque Base EACH1
Tip torque wrench (36 kHz)
Handheld digital tachometer
Reflective tape for the digital tachometer
T connector
Digital Manometer
Waste canister
Prerequisites
Before performing this protocol, record the serial number, calibration due date for all calibrated equipment and other required data in the test datasheet
IT0739A.
CUSA® Clarity System Service and Maintenance Manual
s
Description
Integra Ref. C7000
Integra Ref. C7002
Integra Ref. C7036
Integra Ref. C7414
Integra Ref. C7300EA
Integra Ref. C7005
Integra Ref. C7636
Integra Ref. C7602
Testo 470
RS Component P/N: 845-9731
RS Component P/N: 795-405
Fisher 3461 or Fluke PV350 or Equivalent
Bemis BEH484410 or equivalent
Page 109 of 117