3. Remove the aspiration tubing from the aspiration port at the base of the handpiece.
4. Remove the irrigation tubing from the fitting on the handpiece flue.
5. Slide the flue off the nosecone base along the tip.
CUSA® Clarity System Service and Maintenance Manual
Page 38 of 117
0645403-2-EN