10.5 Handpiece

Dimensions
Length
Diameter
Weight
Cable Length
Frequency
High

10.6 Electrical Requirements

10.6.1 Power Input and Fusing

Voltage
Frequency
Power
Phases
Class
Fuses
NOTE
Only the exact fuse type and rating is to be used. Littelfuse 0215005.HXP, 2 fuses required per console.

10.6.2 Power Cords

United States and Canada:
Meets hospital grade
Cord
Console Connector
Plug
Other Countries:
Cord
Console Connector
CUSA® Clarity System Service and Maintenance Manual
13.03 cm (5.13 in)
2.41 cm (0.95 in)
71 g (0.15 lb)
463.3 cm (182.4 in)
36 kHz
100-240V~
50-60Hz
480VA
1
1
5A, Type T, 1500A IR, 250V, 5mm x 20mm, ceramic body
UL817
3 x 16AWG 5 Meters Color White
C13 IEC 60320-1 V-Lock
NIMA 5-15 Straight
3 x 1mm2 5 Meters Color Black
C13 IEC 60320-1 V-Lock