Tip
The tip is a hollow titanium tube that touches patient tissue. When active, the tip vibrates at an ultrasonic frequency, causing it to fragment tissue.
The tip has two pre-aspiration holes
, one on either side, which help to keep the tip clear and provide better visibility at the surgical site.
Threads
on the end of the tip are for attaching the tip to the connecting body of the handpiece.
The tip is sterile and single use when in a sterile tip pack or can be used up to five (5) times if it is an extended use tip (EUT).
Nosecone
The single-use nosecone attaches to the handpiece, covers the connecting body, and holds the flue in place. It is sterile in a single use tip pack
and nonsterile in an EUT pack.
Integrated O-rings
The o-rings are integrated into the nosecone. They provide stability for the nosecone and prevent fluid leaks into the connecting body.
CUSA® Clarity System Service and Maintenance Manual
Page 19 of 117
0645403-2-EN