Electrical Safety Tests; Required Equipment And Tools - Integra CUSA Clarity System Service And Maintenance Manual

Hide thumbs Also See for CUSA Clarity System:
Table of Contents

Advertisement

• Inspect cart for damage
• Inspect cart wheels for proper rotation and direction
• Inspect the connection attaching the console to the cart.
• Check that all four rubber feet on the console are in place, aligned with the indentations on the cart, and are not damaged.

7.8 Electrical Safety Tests

The electrical safety tests are required to meet the requirements specified in IEC 62353, Edition 2.0 2014-9. Refer to Integra Global Service and Repair
Procedure IT0738 for instructions to complete the tests. This Integra Global Service and Repair Procedure IT0738 and test datasheet are located in
Appendix A.
To reduce the risk to the service technician, perform and pass an electrical test before engaging in any service or maintenance activity on a CUSA
Clarity System.

7.8.1 Required Equipment and Tools

Equipment/Tools
Safety Analyzer
Digital Multi-meter
CUSA® Clarity console mains cable
CUSA® Clarity 36 kHz Handpiece
CUSA® Clarity 36 kHz Curved Ext MicroTip (Single Pack)
CUSA Clarity Quick Connect™ Tubing Set (Single pack)
Waste canister
Jar of water
Contamination guard
CUSA® Clarity footswitch
CUSA® Clarity System Service and Maintenance Manual
Description
Rigel 288 or Rigel 62353 or Fluke ESA620 or equivalent
Keysight 34461A, Fluke 187 or equivalent
Fitted with the console
Integra Ref. C7036
Integra Ref. C7414
Integra Ref. C7300EA
Bemis BEH484410 or equivalent
Any 1-liter jar
Integra Ref. C7005
Integra Ref. C7002
Table 7-3 Functional Test Equipment and Tools