Integra CUSA Clarity System Service And Maintenance Manual page 113

Hide thumbs Also See for CUSA Clarity System:
Table of Contents

Advertisement

Ultrasound Test
Tip Frequency Optimization
Amplitude Test
Irrigation Test
Select Irrigation at 2 ml/min
Press fast flush pedal
Aspiration Test
Set aspiration at 100%
Tested by (Sign and Date): __________________________ dd/mm/yyyy___________
REV
ECO
1
CUSA® Clarity System Service and Maintenance Manual
Result
Expected value
N/A
N/A
> 50%
Result
Expected value
2 to 8 RPM
34 to 47 RPM
Result
Expected value
304 to 436 mmHg
DESCRIPTION
Initial release
Pass
Pass
Pass
Page 113 of 117
Fail
Fail
Fail
UPDATED BY
DATE
Jean-Luc Billard
27 Oct. 2016
0645403-2-EN

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents