Integra CUSA Clarity System Service And Maintenance Manual page 11

Hide thumbs Also See for CUSA Clarity System:
Table of Contents

Advertisement

WARNING
To avoid injury or damage to equipment, place system in standby mode prior to changing tip or clearing a clog.
WARNING
Turn off the power before opening the console - disconnect the power cord.
CAUTION
Federal (USA) law restricts this device to sale by or on the order of a physician.
CAUTION
To avoid product damage, always use the torque base to hold the handpiece while using the torque wrench to tighten or loosen the tip. Avoid over
torqueing the tip.
CAUTION
Only use the cleaning agents listed in the section for cleaning and disinfecting the console. Using solvents or cleaning agents not listed in the
guidelines may damage the plastic exterior of the console.
CAUTION
Do not use sterilization methods involving temperatures in excess of 134ºC (273.2°F) as this may damage the handpiece.
CAUTION
Only an authorized Integra engineer should open the console.
CAUTION
The CUSA Clarity console weight is 29.5 kg (65 lbs). Make sure only individuals who are capable of safely lifting 29.5kg (65 lbs.) lift the console.
CAUTION
Lock all four casters before taking of the console from the cart.
CAUTION
Avoid contact between the CUSA Clarity surgical tip and an activated monopolar surgical instrument.
CAUTION
High frequency surgical equipment can affect the function of medical electrical devices such as ultrasonic aspirators. When using monopolar high
frequency surgical equipment simultaneously with the CUSA Clarity system, contact between the activated monopolar electrosurgery instrument
and the CUSA Clarity surgical tip could affect system functionality. Avoid contact between the CUSA Clarity surgical tip and an activated monopolar
surgical instrument.
CUSA® Clarity System Service and Maintenance Manual
Page 11 of 117