8.5 General Troubleshooting

Troubleshooting
Yes
Is a critical part
damaged?
No
Setup the console with main power
cord, known working handpiece, tubing
set and footswitch.
Turn on the console.
Does power
No
button illuminate
when pressed?
Yes
No
Does the unit power
up? (Fan turning)
Yes
CUSA® Clarity System Service and Maintenance Manual
Call Integra for
assistance.
Check Mains
fuses
Yes
Fuses
Replace Mains
damaged?
fuses
No
Call Integra for
assistance.
Does the
screen light up
Yes
Does the touch
panel respond?
Yes
Does the GUI
report a message in
the Status Bar?
No
Does the GUI
provide a failure
code?
No
Low Aspiration at
the hanpiece tip?
No
No irrigation
flow from the
handpiece
Flue?
No
Excessive misting
at the handpiece tip?
No
No or low
Ultrasonics?
No
End
Figure 8-1
Page 65 of 117
Call Integra for
No
assistance.
Call Integra for
No
assistance.
Explore Status
Yes
Bar root causes
& solutions
Explore Error
Yes
codes matrix
root causes &
solutions
Explore Low
Aspiration at
Yes
the hanpiece
tip root causes
& solutions
Explore No irrigation
Yes
flow from the handpiece
Flue root causes &
solutions
Explore Excessive
Yes
misting at the
handpiece tip
causes & solutions
Explore
Yes
Ultasonics root
causes &
solutions