Integra CUSA Clarity System Service And Maintenance Manual page 106

Hide thumbs Also See for CUSA Clarity System:
Table of Contents

Advertisement

12.0 Measurements of Applied Part Leakage Current
12.1 TN-System (Direct method)
• The Safety Analyzer is connected directly to the mains (TN-System only).
• The DUT (Device Under Test) mains lead is connected to mains socket on the Rigel 288.
• Fit a handpiece, the footswitch, the aspiration and irrigation tubing kit, waste canister, etc. to the console.
• Attached the crocodile clip from the safety tester lead (Applied Part module port 1) to the proximal end of the tip (without the flue).
• From the Rigel 288 main menu, select Manual Mode –> IEC62353 –> AP Leakage Test –> Direct as shown below:
• Press Button
and set the Test Duration to (time) unlimited. Press Button
to exit Settings mode without saving changes.
• Press the green button to allow power on the DUT (Device Under Test) and start measurement.
• Turn on the DUT and wait for the GUI to load.
• Prime the irrigation tubing and perform a handpiece test in order to access the main menu.
• On the DUT main menu, set up the system as follows:
o AMPLITUDE to 100%
o ASPIRATION to 100%
o IRRIGATION to 2ml/min
o TISSUE SELECT to OFF
• Immerse the full tip into a jar feed with 1 liter of water to load the handpiece.
• Press the footswitch Vibrate pedal (orange pedal) fully down, to achieve maximum stroke.
• Read from the Rigel leakage current value.
• Turn off the DUT (Device Under Test) and wait for the system is completely off.
• Normalize measured value according paragraph 7.0 Normalization instructions and record the test result in IT0738A.
CUSA® Clarity System Service and Maintenance Manual
Page 106 of 117
0645403-2-EN