CAUTION
To avoid product damage, always use the torque base to hold the handpiece while using the torque wrench to tighten or loosen the tip. Avoid over
torqueing the tip.
4. Rotate the torque wrench clockwise until it clicks twice.
Note
Keep the torque wrench and base in the sterile field in case you need to change a tip during the procedure and for disassembly after surgery.
Attaching the Nosecone and the Flue
1. Align the white dots by sliding the larger end of the nosecone over the tip and slide it down over the neck of the handpiece.
2. Rotate the nosecone clockwise until it locks into place and the white dot on the nosecone aligns with the while dot on the body of the
handpiece.
Align the white dots on the nosecone and handpiece for proper usage.
Remove stylet from the flue and put it aside in the sterile field if needed to clear the tip during surgery.
CUSA® Clarity System Service and Maintenance Manual
Page 26 of 117
0645403-2-EN