5.5 Attaching the Handpiece
Pass out the handpiece cable connector from the sterile environment, align the red dot on the handpiece cable connector with the red dot on the
handpiece port on the front of the console, and plug in the handpiece cable.
5.6 Connecting the Tubing Set
Note
Do not insert the handpiece cable into the clip at the base of the handpiece.
Note
Properly set and connect the handpiece and contamination guard. See the contamination guard instructions for use for information.
1. Place the canister into the canister holder.
2. Open the cartridge door, insert the cartridge, and slide the cartridge down.
CUSA® Clarity System Service and Maintenance Manual
Page 32 of 117
0645403-2-EN