Chapter 6 - Cleaning and Sterilizing the System

6.1 Overview

Cleaning and sterilization of the CUSA® Clarity system in accordance with the instructions, safety warnings, safety cautions, and user notes listed in
the Operator's Manual is essential to the longevity and proper operations of the system. This chapter provides the trained technician or engineer the
capability to properly clean and sterilize the CUSA Clarity system components and accessories.

6.2 Detaching System Components

After a surgical procedure, keep or discard components and accessories based on the following table.
Keep
Handpiece
Contamination Guard
EUTs (with less than 5 uses)
Torque Base
Foot switch
Power Cord
Sterilization Tray
Table 6-1 Detachable System Components
Detaching the Handpiece
1. Remove the irrigation tubing from the clip at the base of the handpiece.
2. Remove the handpiece cable from the aspiration tubing clips.
CUSA® Clarity System Service and Maintenance Manual
Discard
Nosecone
Flue
Tip Stylet
Torque Wrench
CUSA Quick Connect™ Cartridge and Tubing Set
Aspiration Canister
Irrigation Supply
Specimen Trap (if used)
Single Use Tip
Page 37 of 117
0645403-2-EN