Appendix A: Integra Global Service & Repair Test Procedures (It0738 & It0739) And Test Data Sheets - Integra CUSA Clarity System Service And Maintenance Manual

Hide thumbs Also See for CUSA Clarity System:
Table of Contents

Advertisement

Appendix A:
Integra Global Service & Repair Test Procedures
(IT0738 & IT0739) and Test Data Sheets
1.0 Purpose
This procedure provides a method to conduct CUSA® Clarity Safety test according IEC 62353.
2.0 Scope
This document covers CUSA® Clarity Safety test according IEC 62353.
3.0 Responsibility
Trained personnel shall perform this test.
Global Service & Repair is responsible for updating and maintaining this procedure.
4.0 Reference Documents
CUSA® Clarity Service and Repair Work Instruction (60905375) - Rev A
5.0 Equipment
Equipment
CUSA® Clarity console system (DUT)
Footswitch
CUSA® Clarity 36 kHz Handpiece
CUSA® Clarity 36 kHz Curved Ext MicroTip (Single Pack)
CUSA® Clarity Quick Connect Tubing Set (Single pack)
Contamination guard
CUSA® Clarity 36 kHz Sterilizable Torque Base EACH1
Tip torque wrench (36 kHz)
Waste canister
Jar of water
Digital Multi-meter
Safety Analyzer
CUSA® Clarity System Service and Maintenance Manual
Test Procedure IT0738
CUSA®Clarity GS&R Safety Test Procedure
Issued: 19 Jan 2017
Description
Integra Ref. C7000
Integra Ref. C7002
Integra Ref. C7036
Integra Ref. C7414
Integra Ref. C7300EA
Integra Ref. C7005
Integra Ref. C7636
Integra Ref. C7602
Bemis BEH484410 or equivalent
Any kind of 1 liter
Keysight 34461A, Fluke 187 or equivalent
Rigel 288 or Rigel 62353 or Fluke ESA620 or equivalent
(which have IEC 62353 measurement capability)
Page 102 of 117