Guidance and Manufacturer's Declaration – Electromagnetic Immunity
The CUSA Clarity system is intended for use in the electromagnetic environment specified below.
The customer or the user of the CUSA Clarity system should assure that it is used in such an environment.
IMMUNITY Test
Conducted
RF Immunity
IEC 61000-4-6
in ISM and amateur radio
Radiated
RF Immunity
IEC 61000-4-3
Note
At 80 MHz and 800 MHz, the higher frequency range applies.
Note
These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects,
and people.
CUSA® Clarity System Service and Maintenance Manual
IEC 60601
Compliance Level
Test Level
3 Vrms
150 kHz to 80 MHz
150 kHz to 80 MHz
(3
and 4
ed.)
rd
th
6 Vrms
in ISM and amateur
bands (4
th
ed.)
3 V/m
80 MHz to 2.5 GHz
80 MHz to 2.7 GHz
(3
ed.)
rd
80 MHz to 2.7 GHz
(4
ed.)
th
Electromagnetic Environment
Portable and mobile RF communications equipment
should be used no closer to any part of the system,
including cables, than the recommended separation
distance calculated from the equation applicable to
the frequency of the transmitter.
Recommended separation distance:
3 Vrms
d =1,2P
d =1,2P
6 Vrms
radio bands
d =2,4P
Where P is the maximum output power rating of the
3 V/m
transmitter in watts (W) according to the transmitter
manufacturer and d is the recommended separation
distance in meters (m).
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey
should be less than the compliance level in each
frequency range
Interference may occur in the vicinity of equipment
marked with the following symbol:
Page 94 of 117
- Guidance
80 MHz to 800 MHz
800 MHz to 2.7 GHz
a
b
.
,
0645403-2-EN