Chapter 9 Disassembly And Reassembly Of Sub-Assemblies; Overview; Required Tools And Equipment - Integra CUSA Clarity System Service And Maintenance Manual

Hide thumbs Also See for CUSA Clarity System:
Table of Contents

Advertisement

Chapter 9 - Disassembly and Reassembly of Sub-Assemblies

9.1 Overview

This chapter describes each field replacement part of the system and provides disassembly/ reassembly instructions for their replacement.
This manual describes the proper extent of component disassembly.
WARNING
Turn off the power before opening the console - disconnect the power cord.
CAUTION
Only authorized engineer by Integra can open the console
NOTE
Discard lock washers removed during disassembly and replace with new lock washers when reassembling.
NOTE
When reassembling a lock washer and a flat washer, always place the lock washer between the screw head and the flat washer.

9.2 Required Tools and Equipment

• Torque screwdrivers
o Setting: - 1.62 N-m (14.3 in-lbs) for M4 and smaller
- 2.94 N-m (26.0 in-lbs) for M5 and M6
- 40 N-m (29.5 in-lbs) for M12
• Metric Socket Set up
• M22 flat spanner wrench
• M24 flat spanner wrench
• Metric Allen bits Set
CUSA® Clarity System Service and Maintenance Manual
Page 73 of 117