7.9.2 Required Equipment and Tools

Equipment/Tools
Digital Manometer
CUSA Clarity 36 kHz Handpiece
CUSA Clarity 36 kHz tip
Handheld digital tachometer
Reflective tape for the digital tachometer
CUSA Clarity footswitch
CUSA Clarity Quick Connect Tubing Set (Single pack)
T Connector
Contamination guard
Waste canister

7.9.3 Test Conditions

Perform the electrical safety test at the ambient temperature, humidity, and atmospheric pressure at the testing site.

7.9.4 Test Procedures

Test Setup
• Turn on the console
• Connect the footswitch and handpiece
• Connect a contamination guard and the canister
• Open the cartridge door and stick the reflective tape on the border of the pump head
• Place the cartridge on the pump head, close the door and connect the aspiration tubing set to the canister
• Attach the T connector in line between the distal end of the aspiration tubing set and the handpiece suction port
• Connect the third port of the T connector to the Digital Manometer.
• From the Setup Task menu press the Start Prime button – wait until the prime is completed
Ultrasound Test
• From the Setup Task menu press Start Test button
• When the test is completed, press Details button
• Record that the Tip Frequency Optimization passed the test in IT0739A CUSA Clarity GS&R End User Functional Test Datasheet (Appendix A).
CUSA® Clarity System Service and Maintenance Manual
Description
Fisher 3461 or Fluke PV350 or Equivalent
Integra Ref. C7036
Integra Ref. C7414
Testo 470 or DT2236 or equivalent
RS Components P/N: 845-9731
Integra Ref. C7002
Integra Ref. C7300EA
RS Component P/N: 795-405
Integra Ref. C7005
Bemis BEH484410 or equivalent
Table 7-9 Electrical Safety Test Record Assessment Equipment and Tools
Page 60 of 117