Ultrasonic Root Causes & Solutions
CUSA® Clarity System Service and Maintenance Manual
Little or no vibration
or fragmentation
- Verify that console settings are at
desired levels
- Remove the tip from contact with
Yes
Excessive tip/tissue
tissue, then reactivate vibration.
pressure (stalling tip)?
Note: When first activating
vibration, make sure that there is no
tip/tissue pressure.
No
- Immerse the handpiece tip in sterile
irrigation fluid briefly. This may clear
the blockage.
- Verify that the manifold suction tube connects
Yes
Blocked Tip?
properly to the Handpiece.
- Use the tip cleaner to dislodge the blockage in
the tip. Continue to push the tip cleaner into the
tip until the cleaner becomes visible in the
No
tubing.
Yes
Incorrect flue
installed?
No
- Press the Setup / Standby button to place
the system in Standby mode.
- Remove the handpiece from the
Yes
Tip loose due to
sterile field.
improper assembly?
- Use the sterile torque base and a
disposable sterile wrench to tighten the tip.
(Refer to Cusa Clarity IFU)
No
- Press Start Test button
- Press the Setup / Standby button to place
the system in Standby mode.
- Press Start Test button
Yes
Damaged handpiece?
If the test fail press Detail button and
follow instructions. (Refer to Cusa Clarity
Operator Manual)
- Replace handpiece if problem persists
No
Amplitude level
Yes
fluctuating on the
screen?
No
End
Figure 8-6
Install the correct flue.
Call Integra for assistance