8.10 Error Code Matrix

For all error codes, except those explained below, contact Integra Service and Repair for assistance..
Error Code
Description
E1208
Vacuum Pressure Low
E1220
Irrigation Motor RPM
E1226
Pinch Solenoid Position
E1400
Pinch Solenoid Position
E2201
Broken footswitch microswitch
failure
E2202
Broken footswitch potentiometer
E2203
Broken fast flush footswitch
E2303
Invalid handpiece EEPROM version
E2304
Invalid CRC on handpiece EEPROM
E2305
Handpiece EEPROM parameter out
of range
CUSA® Clarity System Service and Maintenance Manual
Root Cause
Vacuum is lower than expected
Irrigation motor speed is different
from the demand. Check if something
is blocking the irrigation pump head
Pinch valve is not at the expected
position (staying opened)
Pinch valve is not at the expected
position (staying closed)
Microswitch from Ultrasonic
footswitch is defective
Potentiometer from Ultrasonic
footswitch is defective
Microswitch from fast flush footswitch
is defective
Handpiece EEPROM corrupted
Handpiece EEPROM corrupted
Handpiece EEPROM corrupted
Table 8-6
Page 70 of 117
Solution 1
Solution 2
Connect a contamination
Fix leakage on suction
guard
tubing set
Fix blockage at the pump
Call Service for assistance
head
Check for blockage
Call Service for assistance
Check for blockage
Call Service for assistance
Replace footswitch
Call Service for assistance
Replace footswitch
Call Service for assistance
Replace footswitch
Call Service for assistance
Replace handpiece
Call Service for assistance
Replace handpiece
Call Service for assistance
Replace handpiece
Call Service for assistance
0645403-2-EN