Aspiration Troubleshooting Flowchart
No irrigation flow
from the handpiece
Yes
Irrigation pump
not turning?
No
No
IV clamp opened?
No
Is the irrigation tube
No
connected to the handpiece
tip flue?
Yes
Was prime completed
Yes
fully so that fluid filled the
irrigation tubing?
No
CUSA® Clarity System Service and Maintenance Manual
Check on the screen the « On Demand / Constant »
indicator.
- If set to « Constant » the Irrigation pump should
be turning.
- If set to «On Demand » the Irrigation pump
should be turning only when you activate
Ultrasonics.
If these conditions do not exist, call Integra for
assistance.
Open the IV clamp
Connect irrigation tube to
handpiece tip flue
- Return to Setup Task Menu
- Press the Start Prime
button, allowing time for
irrigation fluid to drip from
the tip.
Yes
Irrigation tubing
pinched or kinked?
No
Yes
I.V. bag empty?
No
Irrigation tubing
No
properly sited on
rollers pump
Yes
End
Figure 8-3
Page 68 of 117
Remove the pinch
or kink from the
tubing.
Replace the I.V.
bag.
- Open irrigation pump door
- Place properly irrigation
tubing on the rollers
- Close irrigation pump door
0645403-2-EN