• From the Rigel 288 main menu, select Manual Mode –> IEC62353 –> Equipment Leakage Test –> Direct as shown below:
• Press Button
• Press the green button to allow power on the DUT (Device Under Test) and start measurement.
• Turn on the DUT and wait for the GUI to load.
• Prime the irrigation tubing and perform a handpiece test in order to access the main menu.
• On the DUT main menu, set up the system as follows:
o AMPLITUDE to 100%
o ASPIRATION to 100%
o IRRIGATION to 2ml/min
o TISSUE SELECT to OFF
• Press the footswitch Vibrate pedal (orange pedal) fully down, to achieve maximum stroke.
• Read from the Rigel leakage current value.
• Turn off the DUT (Device Under Test) and wait for the system is completely off.
• Normalize the measured value according paragraph 7.0 Normalization instructions and record the test result in IT0738A.
11.2 IT-System (Alternative method)
• The Safety Analyzer is connected directly to the mains (IT-System only).
• The DUT (Device Under Test) mains lead is connected to mains socket on the Rigel 288.
• Connect the handpiece and the footswitch.
• Attached the crocodile clip from the safety tester lead (Applied Part module port 1) to the proximal end of the tip (without the flue).
• Attached the test lead from the safety tester (Applied Part module port 2) to the Equipotential Earth Ground pin of the console.
• From the Rigel 288 main menu, select Manual Mode –> IEC62353 –> Equipment Leakage Test –> Alternative as shown below:
• Execute the EQUIPMENT LEAKAGE CURRENT measurement by pressing green button.
• A Warning message as described in paragraph 8.0 Earthing system appears on the Rigel 288 screen to inform the user that the test is
performed under IT-System network. Press the green button again to perform the test.
• Normalize measured value according paragraph 7.0 Normalization instructions and record the test result in IT0738A.
CUSA® Clarity System Service and Maintenance Manual
and set the Test Duration to (time) unlimited. Press Button
to exit Settings mode without saving changes.