Integra CUSA Clarity System Service And Maintenance Manual page 47

Hide thumbs Also See for CUSA Clarity System:
Table of Contents

Advertisement

EUT Sterilization Tray Configuration
1. Place the handpiece in the appropriate slot
2. Wrap the handpiece cable around the posts
3. Place the handpiece connector in the appropriate slot
4. Optional. Place the nosecone in the appropriate slot
5. Optional. Place the flue
anywhere on the bottom of the tray.
6. Place the torque base in the appropriate slot
7. Optional. Place the Extended Use Tips (up to 3) in the appropriate slots
8. Put on the lid
on the sterilizer tray.
9. Close and latch the lid.
Sterilization Parameters
Sterilize the CUSA Clarity handpiece and components with steam.
Sterilization Tray Packaging
Packaged in a sterilized tray, the handpiece and components are:
• Wrapped—Sterilization tray double wrapped in hospital CSR material
• Flash (Unwrapped)—Sterilization tray unwrapped
The cycles below are acceptable for sterilizing the CUSA Clarity handpiece and components
Packaging
Temp
132°C/269.6°F
134°C/273.2°F
Wrapped
134°C/273.2°F
Flash
132°C/269.6°F
CUSA® Clarity System Service and Maintenance Manual
and align it with the outline on the bottom of the tray.
.
and align it with the outline on the bottom of the tray.
and align it with the outline on the bottom of the tray.
and align it with the outline on the bottom of the tray.
and align them with the outline on the bottom of the tray.
Type
Prevac
Prevac
Prevac
Prevac
Table 6-3 Sterilization Cycles
Page 47 of 117
Time
Dry Cycle
4 min
30 min
3 min
30 min
18 min
30 min
4 min
None
0645403-2-EN