EMS AIR-FLOW MASTER PIEZON Operation Instructions Manual page 135

Hide thumbs Also See for AIR-FLOW MASTER PIEZON:
Table of Contents

Advertisement

No water jet coming from the Piezon
handpiece
No powder/air jet coming from the
$,5)/2: RU 3(5,2)/2: handpiece
Water leakage between the handpiece
cord and the handpiece
:DWHU OHDNDJH EHWZHHQ WKH 3(5,2)/2:
handpiece and the nozzle
Air and powder leakage on the main unit
Air and powder is leaking at powder
chamber
Handpiece $,5)/2: RU 3(5,2)/2:
HI¿FLHQF\ GHFUHDVHV
/HDNDJH RQ WKH XQLW ZDWHU VXSSO\
&KHFN WKH ZDWHU VXSSO\ DQG ZDWHU ¿OWHU
Check the water adjustment of your unit
Check the connections between the unit, the cord and the handpiece
Check for presence of water at the cord outlet port
Send the handpiece to an authorized EMS Repair Center
&KHFN WKH FRPSUHVVHG DLU VXSSO\ DQG DLU ¿OWHU
Check for presence of powder/air at cord outlet port
Eliminate residual powder present in the channels using the cleaning needles provided
Check the quantity of powder in your unit
Send the system to an authorized EMS Repair Center
Check the condition of the handpiece connecting O-rings
Send the system and the handpiece cord to an authorized EMS Repair Center
Verify the position of the nozzle and its position on the handpiece
Verify that no dirt remains between the nozzle and the handpiece
9HULI\ WKDW QRWKLQJ DIIHFWV WKH DOLJQPHQW RU SUHYHQWV ¿[LQJ WKH QR]]OH
Replace the nozzle
Replace the handpiece
6HQG WKH V\VWHP DQG WKH KDQGSLHFH FRUG WR DQ FHUWL¿HG (06 5HSDLU &HQWHU
Verify the positioning of the bottles
Check the O-rings under the silicone caps of the powder chamber holders
Verify the cap of the powder chamber and the O-ring positioning
Verify the check valve and the O-ring under the powder chamber
Remove all powder dust between the bottle holder and the powder chamber
Send the unit to an authorized EMS Repair Center
Check the seal condition and the cleanliness of the thread of the powder chamber cap
Replace the seals if necessary
Clean the sealing surfaces carefully
Replace the powder chambers
Send the unit to an authorized EMS Repair Center
Verify the quantity of powder in your unit
Eliminate residual powder present in the channels using a cleaning needle
6HQG WKH V\VWHP DQG WKH KDQGSLHFH FRUG WR DQ FHUWL¿HG (06 5HSDLU &HQWHU
Check the water supply connection and inlet pressure
Send the unit to an authorized EMS Repair Center
117

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents