3.5.4 Conclusions

Carry out all preliminary checks and make sure that they were completed successfully
before connecting the patient to the Falco 202 Evo lung ventilator.
CAUTION - Preliminary checks phase failed.
Please see Alarms chapter and/or Trouble Shooting chapter.
Please contact the nearest Siare Support Centre or any other support
centres authorised by Siare.
The intensive care lung ventilator must be inspected and serviced once it
reaches 1000 hours of operation or, in case of limited use, at least once
every 6 months.
All maintenance and/or repair interventions require full knowledge of the lung
ventilator, and therefore such operations must be carried out only by highly
qualified staff, specifically trained and authorised by SIARE.
Any improper intervention or unauthorised modification may affect the
device's safety, putting the patient at risk.
Falco 202 Evo
During ventilation, set the main power supply switch to OFF ( 0 ).
The system activates the main power failure alarm: silence the
alarm.
Restore the main switch and set it to ON ( I ).
Close the medical gas supply during lung ventilator operation.
The system activates the gas supply failure alarm: silence the
alarm.
Restore the medical gas supply.
WARNING !! Patient injury hazard
Check the alarm limits setup values before connecting a patient to the
lung ventilator.
Change the alarm limits setup based on the clinical situation.
WARNING !! User and patient injury hazard
3-35