GE LOGIQ 200 Quality Assurance Manual page 167

Table of Contents

Advertisement

GE MEDICAL SYSTEMS
REV 0
2--3--2
System Diagnosis (Continued)
TEST NAME: B MODE CINE TEST
PURPOSE
To check B--mode CINE.
CRITERIA
Refer to result shown on the monitor.
Action if the test fails
Do followings as many as the number of frames.
--Set the cine numbers.
--Unfreeze the current image.
--Freeze the current image.
--Read Top, middle, and bottom horizontal line of B--mode image.
--Compare each horizontal line.
The following Assemblies are necessary for this test
Probe, RBF, MVP, MSTE, SCAN, Monitor
MVP
FRONT_END
From RBF
To NOVA_BUS
Buffer
Image
Frame
Softener
Average
PEAK
HOLD
B/M
Front_end
Control
DSP BLOCK
USC BUS
Buffer
I/F
LOGIQ 200 PRO Series PROPRIETARY MANUAL
B MODE
ITP
ALM
CRDTF
XYADDR
Gen.
CINE
CINE
CINE
Control
Memory
SAP
FIFO
DSP
2--30
POST
Image
Storage RAM
Image
Display Control
Video
XY
Lut
memory
Interlace
ZOOMER
Convert
Non--Int
Convert
Video
Decoder
Sync
Generator
XY
Memory
Program
Memory
DIAGNOSTICS
2242594
Video
DAC
Video
DAC