Gas Sterilization
Perform the probe cleaning instructions before following the gas sterilization procedures.
Ethylene oxide or formalin gas sterilization process may be used on surgical and biopsy probes
only.
Follow the instructions, recommendations and precautions provided by the manufacturer of the
sterilization equipment and gas supplier.
NOTE: Do not use process cycles that exceed 60 degrees C or pressures that vary from
normal atmosphere.
Ultrasound transducers are easily damaged by improper handling, storage, or
cleaning. Failure to follow OEM Specifications and Precautions can result in
equipment damage and replacement expense liability. Never immerse the transducer
connector or probe adapters into any liquid. Refer to the illustrations in the system
Operator Manual. Avoid mechanical shock or impact to the transducer. Do not apply
excessive bending or pulling force to the cable.
ULTRASOUND QUALITY ASSURANCE REFERENCE MANUAL
CAUTION
1-3
2262684-100