GE LOGIQ 200 Quality Assurance Manual page 116

Table of Contents

Advertisement

GE MEDICAL SYSTEMS
REV 0
2–3–2
System Diagnosis (Continued)
TEST NAME
SYSTEM TEST
BEAM FORMER TEST
HV TEST
ADC TEST
DSC FRONT/END TEST
B Mode Function ALM test
B Mode Interpolation test
B Mode Coordination test
B Mode X–Y Memory test
B Mode CINE test
M Mode Function TEST
MST TEST
SYSTEM DRAM TEST
KEYBOARD TEST
PIOP TEST
MONITOR TEST
POWER SUPPLY TEST
ROM VERSION TEST
TABLE 2–2
RESTRICTED CONDITIONS FOR SYSTEM DIAGNOSIS TESTS
FUNCTION OR SETTING PARAMETERS
Console parameters are changed automatically.
Display Format= Linear Depth= 4Cm Softener= Off
Frame average= Off Frame rate= High B gain= 60
TGC= 0 Skin Depth= 0
Console parameters are not changed
Display Format= Linear Depth= 10Cm Softener= Off
Frame average= Off
Display Format= Convex Softener= Off
Display Format= Linear
Display Format= Linear
Display Format= Convex
Display Format= Convex
Display Format= Linear
Display Format= Linear Sweep Speed= High
Mode = Full M mode
Console parameters are not changed
Console parameters are not changed
Console parameters are not changed
Console parameters are not changed
Console parameters are not changed
Console parameters are not changed
Console parameters are not changed
2–15
a
LOGIQ
200 PROPRIETARY MANUAL
2138854
DIAGNOSTICS