GE LOGIQ 200 Quality Assurance Manual page 156

Table of Contents

Advertisement

GE MEDICAL SYSTEMS
REV 0
2--3--2
System Diagnosis (Continued)
TEST NAME: SYSTEM TEST
PURPOSE
To check all the system functions sequentially except Beam former test, ADC test, keyboard test,
Monitor test and ROM version.
CRITERIA
Refer to result shown on the monitor.
DISPLAY FORMAT OF RESULTS
Refer to result shown on the monitor as in ILLUSTRATION 2--8.
Service Utility MENU :
Current Software Version is 3.0
NAME
*System test
Scan Test
TBF Test
Linear/Convex Path
Sector Probe Path
BPHV test
MVP Hardware Test
B mode Function
B mode Cine
M mode Function
MSTE test
Keyboard
PIOP
Monitor
Power Supply
Exit
Press 'ctrl'+'X'to Exit
Action if the test fails
Refer to fail code listed in TABLE 2--2 and take actions.
The following Assemblies are necessary for this test
BPHV, PROBE, RBF, TBF, SCAN, MVP, MSTE, Monitor
LOGIQ 200 PRO Series PROPRIETARY MANUAL
System Diagnosis
Time
21 M
0 M
0 M
0 M
0 M
1 M
1 M
1 M
8 M
2 M
4 M
0 M
1 M
0 M
1 M
SYSTEM TEST MENU--RESULT
ILLUSTRATION 2--8
2--19
99/03/12 09:16:37
Status
Passed
Checked
Checked
Checked
Checked
Checked
Running
Passed
Fail:###
Checked
No PIOP
Checked
2242594
DIAGNOSTICS