GE LOGIQ 200 Quality Assurance Manual page 193

Table of Contents

Advertisement

With gains set to obtain a uniform phantom image, freeze the image. Complete the questions regarding the uniformity of the image by marking the appropriate box using this key.
1) Agree
Transducer #1
1) The average brightness at edge of the scan is the same as the
average brightness in the middle.
q 1
2) There are no vertically or radially oriented shadows from array element
dropout.
q 1
3) There are no brightness transitions between focal zones.
q 1
Transducer #1
Manufacturer of Phantom: _____________________ Model: __________
Object Depth
(Look at object's depth over a wide a range as possible to cover
maximum range of the scanner)
1
2
3
Object Size
1
2
3
Object Contrast
1
2
3
Detectability
1
2
3
Complete
Complete
Complete
Partial
Partial
Partial
Not Seen
Not Seen
Not Seen
Changed for Worse
1
2
3
NO
NO
With system sensitivity set up for visualizing echogenicity as deeply as possible into the phantom, what is the maximum depth you can visualize the background
echographic pattern?
Transducer #1
q
q
Less than 3 cm
6 cm
q
q
3cm
6.5 cm
q
q
3.5 cm
7 cm
q
q
4 cm
7.5 cm
q
q
4.5 cm
8 cm
q
q
5 cm
8.5 cm
q
q
5.5 cm
9 cm
2) Disagree, slight non uniformities present
q 2
q 3
q 2
q 3
q 2
q 3
LOW CONTRAST OBJECTIVE DETECTABILITY (Optional)
Complete only if you use a phantom
4
5
4
5
4
5
4
5
Complete
Complete
Partial
Partial
Not Seen
Not Seen
4
5
NO
NO
NO
Mark the appropriate box.
q
q
9.5 cm
13 cm
q
q
10 cm
13.5 cm
q
q
10.5 cm
14 cm
q
q
11 cm
15 cm
q
q
11.5 cm
16 cm
q
12 cm
q
12.5 cm
UNIFORMITY
1) The average brightness at edge of the scan is the same as the
average brightness in the middle.
q 1
2) There are no vertically or radially oriented shadows from array element
dropout.
q 1
3) There are no brightness transitions between focal zones.
q 1
Manufacturer of Phantom: _____________________ Model: __________
Object Depth
(Look at object's depth over a wide a range as possible to cover
maximum range of the scanner)
1
2
Object Size
1
2
Object Contrast
1
2
Detectability
1
2
Complete
Complete
Partial
Partial
Not Seen
Not Seen
Changed for Worse
1
2
NO
NO
PENETRATION
q
q
Less than 3 cm
q
q
3cm
q
q
3.5 cm
q
q
4 cm
q
q
4.5 cm
q
q
5 cm
q
q
5.5 cm
3) Disagree, major non uniformities present
Transducer #2
q 2
q 3
q 2
q 3
q 2
q 3
Transducer #2
3
4
3
4
3
4
3
4
Complete
Complete
Partial
Partial
Not Seen
Not Seen
3
4
NO
NO
Transducer #2
q
6 cm
9.5 cm
q
6.5 cm
10 cm
q
7 cm
10.5 cm
q
7.5 cm
11 cm
q
8 cm
11.5 cm
q
8.5 cm
12 cm
q
9 cm
12.5 cm
5
5
5
5
Complete
Partial
Not Seen
5
NO
q
13 cm
q
13.5 cm
q
14 cm
q
15 cm
q
16 cm

Hide quick links: